Varian VHS-6 Instruction Manual Page 14

  • Download
  • Add to my manuals
  • Print
  • Page
    / 30
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 13
1-1
SECTION I
INSTALLATION
Before unpacking and installing the VHS-6 Diffusion Pump, the user should thoroughly familiarize
himself with this instruction manual and the diffusion pump specifications (see Table 1). He should
also examine all other technical material supplied in order to gain a better understanding of the oper-
ating principles, limitations, correct application, and the hazards involved with the use of this equip-
ment.
1-1 SPECIFICATIONS
Table 1 lists the specifications for all standard type 0184 models of the pump.
Table 1. Pump Specifications
Power Rating 2200 watts
Optimum Operating Range 1 x 10
-3
to <5 x 10
-9
(Torr)
.133 Pa to <.000000665 Pa
Maximum Pumping Speed
(liters/sec – air) 2400 with std cold cap; 1600 with extended cold cap
(liters/sec – helium) 3000 with std cold cap; 2000 with extended cold cap
Maximum Throughput 3.5
(Torr liters/sec at 0.01 Torr (13.3 Pa))
Maximum Forepressure No Load: 0.65 Torr (86.45 Pa)
Full Load: 0.55 Torr (73.15 Pa)
Backstreaming Rate at Pump Inlet <5 x 10
-4
mg/cm
2
/minute (with std cold cap)
Power Required (approximately) 120, 240, 208, 50/60 Hz, single phase
Warmup Time 9 minutes
Cooldown Time (using quick cool coil) 10 minutes
Fluid Charge 500 cc
Cooling Water Requirements
(max. inlet temperature) 60/80 degrees F
(max. outlet temperature at foreline) 120 degrees F
(general flow rate) 0.25 gpm
Pressure drop across coils 15 psi
Backing Pump Size Recommended 17 cfm for maximum throughput
Jet Assembly 4-stage, self-aligning, stainless steel
Foreline Baffle Stacked half moon
Cold Cap Nickel-plated Copper
Water Connections
1
8 FPT
Thermal Switches Manual reset at 300°F
Page view 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 29 30

Comments to this Manuals